Advanced instruments inc, Accuracy overview – Analytical Industries GPR-1600 Series Trace PPB Oxygen Analyzer User Manual
Page 13

Advanced Instruments Inc.
 
 
 
13
Accuracy Overview
 
Single Point Calibration: As previously described 
the galvanic oxygen sensor generates an electrical 
current proportional to the oxygen concentration in 
the sample gas. In the absence of oxygen the sensor 
exhibits an absolute zero, e.g. the sensor does not 
generate a current output in the absence of oxygen. 
Given these linearity and absolute zero properties, 
single point calibration is possible. 
 
Pressure: Because sensors are sensitive to the 
partial pressure of oxygen in the sample gas their 
output is a function of the number of molecules of 
oxygen 'per unit volume'. Readouts in percent are 
permissible only when the total pressure of the 
sample gas being analyzed remains constant. The 
pressure of the sample gas and that of the calibration 
gas(es) must be the same (reality < 1-2 psi). 
 
Temperature: The rate oxygen molecules diffuse into the sensor is controlled by a Teflon membrane otherwise known as an 
'oxygen diffusion limiting barrier' and all diffusion processes are temperature sensitive, the fact the sensor's electrical output 
will vary with temperature is normal. This variation is relatively constant 2.5% per ºC. A temperature compensation circuit 
employing a thermistor offsets this effect with an accuracy of +5% or better and generates an output function that is 
independent of temperature. There is no error if the calibration and sampling are performed at the same temperature or if the 
measurement is made immediately after calibration. Lastly, small temperature variations of 10-15º produce < 1% error. 
 
Accuracy:
In light of the above parameters, the overall accuracy of an analyzer is affected by two types of errors:
 
1) those producing 'percent of reading errors', illustrated by Graph A below, such as +5% temperature compensation
circuit,
tolerances of range resistors and the 'play' in the potentiometer used to make span adjustments and 
 
2) those producing 'percent of full scale errors', illustrated by Graph B, such as +1-2% linearity errors in readout devices, which 
are really minimal due to today's technology and the fact that other errors are 'spanned out' during calibration. 
 
Graph C illustrates these 'worse case' specifications that are typically used to develop an analyzer's overall accuracy statement 
of < 1% of full scale at constant temperature or < 5% over the operating temperature range. QC testing is typically < 0.5% 
prior to shipment. 
 
 
 
 
 
 
 
 
 
 
 
 
Example 1: As illustrated by Graph A any error, play in the multi-turn span pot or the temperature compensation circuit, 
during a span adjustment at 20.9% (air) of full scale range would be multiplied by a factor of 4.78 (100/20.9) if used for 
measurements of 95-100% oxygen concentrations. Conversely, an error during a span adjustment at 100% of full scale range is 
reduced proportionately for measurements of lower oxygen concentrations. 
 
 
Refer to the Calibration section for additional details. 
 
