Auto Range Sampling 
The display will shift to the next higher range when the oxygen reading exceeds 99.9% of the current range. The display will 
shift to the next lower range when the oxygen reading drops to 85% of the next lower range. 
 
For example, if the analyzer is reading 1 ppm on the 0-10 ppm range and an upset occurs, the display will shift to the 0-100 
ppm range when the oxygen reading exceeds 9.99 ppm. Conversely, once the upset condition is corrected, the display will shift 
back to the 0-10 ppm range when the oxygen reading drops to 8.5 ppm. 
 
Procedure:
From the SAMPLE menu, advance the cursor (*) to the “Auto Range” option and press ENTER:
SAMPLE
* Auto Range 
 Manual Range 
 Bypass 
 Standby 
 
 
Standby
Auto Range 
85⁰F 100Kpa 
 
12/31/07 12:00:00 
Note:
To provide for the possibility of an optional automated sample system, the system displays the message “Opening 
Sample Valves” at this time. This message does not apply to analyzers equipped with the standard manually operated 
sample system. 
Similarly, the Bypass and Standby options do not apply to analyzers equipped with manual sample systems.
 
Within seconds the system assesses the oxygen concentration, selects the appropriate range (as described above) and returns 
to the MAIN MENU in the “Sample” mode. On the top line at the bottom of the menu, the Auto Range mode is indicated as is 
the current full scale range. 
 
* MAIN MENU
 Sample 
 Span 
 Zero 
 Alarm 
 System 
 Standby 
 
Sample
5.00 PPM
Auto Range 
85⁰F 100Kpa 
0 to 10 PPM 
12/31/07 12:00:00 
 
 
 
 
 
 
 
 
 
 