2 quality control certification, Advanced instruments inc – Analytical Industries GPR-1600 Series Trace PPB Oxygen Analyzer User Manual
Page 3
Advanced Instruments Inc.
3
2 Quality Control Certification
Date:
Customer: Order No.:
Pass
Model:
GPR-1600 ppm Oxygen Analyzer S/N _____________________
Sensor:
( ) GPR-12-333 ppm Oxygen Sensor
( ) XLT-12-333 ppm Oxygen Sensor S/N _____________________
Accessories:
Owner’s Manual
CABL-1008 Power Cord
TOOL-1001 5/16” Combination Wrench
Configuration:
Ranges: 0-1 ppm, 0-10 ppm, 0-100 ppm, 0-1000 ppm
A-1146-10 PCB Assembly Main / Display Software V. ______
( ) A-1174-10 PCB Assy Power Supply / Interconnect, 4-20mA Range ID
( ) A-1174-10C PCB Assy Power Supply / Interconnect, 5x Relay Contacts Range ID
( ) Stainless steel sensor housing, manual flow control and bypass valves, ¼” compression
type fittings for sample inlet and vent
( ) Delete bypass valve from above (T and TO options)
( ) Sample, span, zero inlet solenoid valves
( ) Temperature controlled heater system 85°F specify: ( ) 110VAC ( ) 220VAC
Power: 100/120/220/250 VAC (universal without temperature controlled heater systems)
Enclosure: ( ) Std. panel mount (“T”) 7.5x10.8x12”; ( ) “TO” option 7.75x 7.75x12”
( ) Bezel for 19” rack mmount 19x12x12” option
( ) GPR-1600-W option general purpose wall mount 12x12x8”
( ) GPR-1600-W306 option general purpose panel mount 18.2x16x10”
Test
System start-up diagnostics satisfactory
Auto/manual range
Alarm relays activate/deactivate with changes in O
2
concentration
Alarm bypass
Analog outputs: Signal output 4-20mA
Range ID: ( ) 4-20mA or ( ) 5x relay contacts plus 1x common
Recovery from air to < 10 ppm in < 15 minutes
Baseline drift on zero gas <
± 2% FS over 24 hour period
Noise level <
± 1.0% FS
Span calibration gas value
Span adjustment within 10-50% FS
Peak to peak over/under shoot < 0.5% FS
Overall inspection for physical defects
Options
Label analyzer “Oxygen Service” in accordance with P-1507 Rev-1; see certificate next page.
Notes