ZERO CALIBRATION

IN PROGRESS . . .

5.00 PPM


After the oxygen reading has stabilized, press ENTER to complete the Zero Calibration. If the user attempts to initiate the ZERO
CALIBRATION function while the oxygen reading is above 50% of the most sensitive or lowest range, the system displays the
message “CALIBRATION FAILED” and returns to the “Sample” mode.