Advanced instruments inc – Analytical Industries GPR-1600 Series Trace PPB Oxygen Analyzer User Manual
Page 29

Advanced Instruments Inc.
 
 
 
29
ALARM Sample
* Set Alarm 1 
 Set Alarm 2 
 Alarm 1 HI 
 Alarm 2 HI 
 Alarm 1 ON 
 Alarm 2 ON 
 Alarm Timeout 
 
 
Auto Range 0 to 10 PPM 
85⁰F 100Kpa 12/31/07 12:00:00 
 
Advance the cursor (*) to the “Set Alarm 1” option and press the green ENTER key to accept the selection. 
The following menu appears: 
 
Sample
20%
 Press UP or DOWN 
 to change value 
 ENTER to Save 
 ESC to Return 
 
 
 
 
Auto Range 0 to 10 PPM 
85⁰F 100Kpa 12/31/07 12:00:00 
 
Follow the prompt above and press the ENTER key to save the alarm value (% full scale) or ESC to return to the MAIN MENU. 
Within a few seconds after pressing the ENTER key, the system returns to the MAIN MENU. 
 
Repeat the above steps for “Set Alarm 2”. 
 
Configure Alarm 1 and Alarm 2 by advancing the cursor (*) to the desired feature as illustrated below. 
 
ALARM Sample
 Set Alarm 1 
 Set Alarm 2 
* Alarm 1 HI / LO 
* Alarm 2 HI / LO 
* Alarm 1 ON / OFF 
* Alarm 2 ON / OFF 
 Alarm Timeout 
Auto Range 0 to 10 PPM 
85⁰F 100Kpa 12/31/07 12:00:00 
 
Press the ENTER key to toggle between the settings: HI and LO and/or ON and OFF. 
After toggling, the system returns to the MAIN MENU. To confirm selection, re-access the ALARM menu above. 
Advance the cursor (*) to the “Alarm” option and press the green ENTER key to accept the selection. 
 