Advanced instruments inc, Calibration – Analytical Industries GPR-1600 Series Trace PPB Oxygen Analyzer User Manual
Page 33

Advanced Instruments Inc.
 
 
 
33
Calibration
 
The electrochemical oxygen sensors manufactured by Analytical Industries Inc. (dba Advanced Instruments) generate an 
electrical current that is linear or proportional to the oxygen concentration in the sample gas. In the absence of oxygen the 
sensor exhibits an absolute zero, e.g. the sensor does not generate a current output in the absence of oxygen. Given the 
properties of linearity and an absolute zero, single point calibration is possible. 
 
As described below, zero calibration is recommended only when the application (or user) demands optimum accuracy for 
analysis below 5% of the most sensitive or lowest range available on the analyzer. Span calibration in one of the forms 
described below is sufficient for all other measurements. When employed zero calibration should precede span calibration. 
 
Zero Calibration 
Despite the absolute zero inherent in electrochemical oxygen sensors, the reality is that analyzers display an oxygen reading 
when sampling a zero gas due to: 
 
¾ Contamination or quality of the zero gas 
¾ Minor leakage in the sample line connections 
¾ Residual oxygen dissolved in the sensor’s electrolyte 
¾ Tolerances of the electronic components 
 
The zero capability (low end sensitivity) of every analyzer is qualified prior to shipment. However, because the factory sample 
system conditions differ from that of the user, no ZERO OFFSET adjustment is made to the analyzer by the factory 
 
Span Calibration 
Involves periodically, see Intervals section below, checking and/or adjusting the electronics to the sensor’s signal output at a 
given oxygen standard or span. Maximum drift from calibration temperature is approximately 0.11% of reading per °C. The 
frequency of calibration varies with the application conditions (potential for contamination), the degree of accuracy required by 
the application and the quality requirements of the user. However, the interval between span calibrations should not exceed 
three (3) months. 
 
Note: Regardless of the oxygen concentration of the standard used, the span calibration process takes approximately 10 
minutes, however, the time required to bring a ppm analyzer back on-line can vary, see Online Recovery Time below. 
 
Considerations 
When it comes to the calibration of oxygen analyzers utilizing an electrochemical oxygen sensor circumstances vary widely from 
the ideal conditions that exist at the factory to a variety of differing circumstances users encounter in the field. The following 
describes the most common factors and reasons that they influence the calibration procedures. 
 
Factor Reasons 
 
Intervals: 
 
All electrochemical sensor based analyzers require periodic, e.g. weekly intervals to a 3 
month maximum, calibration to ensure accuracy and ascertain whether the sensor has been 
contaminated or otherwise damaged while in service. 
 
Conditions: 
 
Calibrate at the temperature and pressure of the sample. 
 
Analysis Level Required: 
 
Continuous analysis below 5% of the most sensitive or lowest range available: 
ZERO CALIBRATION followed by SPAN CALIBRATION with good quality gases is 
recommended (for optimum accuracy) when: 
- the analyzer and/or O2 sensor is initially installed, 
- the sample system connections are modified, 
- the O2 sensor is replaced. 
 
Note: It is not necessary to repeat the ZERO CALIBRATION with subsequent periodic 
 SPAN CALIBRATION unless desired or one of the above events occurs. 
 
All other analysis: SPAN CALIBRATION is sufficient. Procedure varies with factors. 
 
