OM17-P4000VIS-NIR

6-8

Figure 15

Make sure the o-ring is pressed in the groove all the way before installing the window

sheath. Now secure the window sheath using the four hex screws. Tighten down the

hex screw on the side of the window housing. Move up to the top and connect the EC

point wires to the terminal block. Be careful with the fiber optic cable when putting the

front cover back on. It can be pinched between cover and probe. Once cover is on

secure with the bolts.