Veris Technologies VIS_NIR Spectrophotometer Probe - Operating Instructions User Manual
Page 58
OM17-P4000VIS-NIR
6-8
Figure 15
Make sure the o-ring is pressed in the groove all the way before installing the window
sheath. Now secure the window sheath using the four hex screws. Tighten down the
hex screw on the side of the window housing. Move up to the top and connect the EC
point wires to the terminal block. Be careful with the fiber optic cable when putting the
front cover back on. It can be pinched between cover and probe. Once cover is on
secure with the bolts.