CARE AND MAINTENANCE

CARE AND MAINTENANCE

CARE AND MAINTENANCE

CARE AND MAINTENANCE

CARE AND MAINTENANCE

The RS232 Interface Accessory once installed does not require maintenance.

TROUBLESHOOTING

TROUBLESHOOTING

TROUBLESHOOTING

TROUBLESHOOTING

TROUBLESHOOTING

PROBABLE CAUSE(S)

Print menu settings not
properly set up.

Cable connections.

SYMPTOM

RS232 interface not
working.

REMEDY

Verify interface settings in
Print menu correspond to
those of peripheral device.

Check cable connections.

Error Codes

Error Codes

Error Codes

Error Codes

Error Codes

When the RS232 interface is installed in Precision Standard and Analytical Standard
balances, the following error codes are added to the standard list of error codes:

4.0

Bad RS232 frame. Check RS232 menu parameters and correct.

4.4

RS232 buffer is full. May occur if no printer or computer is connected to the
interface. To clear buffer turn balance off, or enter and exit Print menu and select
END.

4.5

Function is disabled by the Lockswitch.

SERVICE INFORMATION

SERVICE INFORMATION

SERVICE INFORMATION

SERVICE INFORMATION

SERVICE INFORMATION

If the Troubleshooting section does not resolve or describe your problem, you will need
to contact an authorized Ohaus Service Agent. For Service assistance in the United
States, please call Ohaus Corporation toll-free at (800) 526-0659. An Ohaus Product
Service Specialist will be available to help you.