beautypg.com

Calibration menu, Calibration menu protection, Calibration masses – Ohaus AS Series Electronic Balances User Manual

Page 47: Calibration weights

background image

13

CALIBRATION MENU

CALIBRATION MENU

CALIBRATION MENU

CALIBRATION MENU

CALIBRATION MENU

The Model AS60C Analytical Standard balance can be calibrated in two ways: Span
calibration or Linearity calibration. Span calibration resets the balance’s weighing
range using two weight values: zero and a weight value at or near the balance’s
capacity. Linearity calibration minimizes deviation between actual and displayed
weights within the balance’s weighing range. Three weight values are used: zero, a
weight value within the balances weighing range, and a weight value at or near the
balance’s specified capacity. The following table shows the sequence in which
submenus appear on the Calibration menu.

CALIBRATION MENU TABLE

CALIBRATION MENU TABLE

CALIBRATION MENU TABLE

CALIBRATION MENU TABLE

CALIBRATION MENU TABLE

Selects span calibration.

Selects linearity calibration.

Used to exit the Calibration menu.

Calibration Menu Protection

Calibration Menu Protection

Calibration Menu Protection

Calibration Menu Protection

Calibration Menu Protection

Calibration may be locked out to prevent
unauthorized personnel from changing
calibration. To lock out calibration menu,
refer to the section titled Menu Lock-Out
Protection.

NOTE

NOTE

NOTE

NOTE

NOTE: If calibration has been locked out,
you will not be able to access it.

Calibration Masses

Calibration Masses

Calibration Masses

Calibration Masses

Calibration Masses

Before beginning calibration, make sure
masses are on hand. If you begin calibra-
tion and realize masses are not available,
either turn the balance off, or go through
the procedure without masses. The bal-
ance will use previously stored calibration
data. Calibration should be performed as
necessary to ensure accurate weighing.
Masses required to perform the proce-
dures are listed in the adjacent table.

CALIBRATION MASSES

CALIBRATION MASSES

CALIBRATION MASSES

CALIBRATION MASSES

CALIBRATION MASSES

LINEARITY

LINEARITY

LINEARITY

LINEARITY

LINEARITY

SPAN ONLY

SPAN ONLY

SPAN ONLY

SPAN ONLY

SPAN ONLY

MODEL

MODEL

MODEL

MODEL

MODEL

MASS

MASS

MASS

MASS

MASS

MASS

MASS

MASS

MASS

MASS

AS60C 20g, 50g 50g

Masses must meet or exceed ASTM Class 1
Tolerance. Calibration masses are available
as accessories.