30

Error Codes

Error Codes

Error Codes

Error Codes

Error Codes

The following list describes the various error codes and which can appear on the
display and the suggested remedy.

Data Errors

Data Errors

Data Errors

Data Errors

Data Errors

0.0

Transient error (hardware error, probably static discharge). If error persists,
the balance must be serviced.

Zero/Tare Errors

Zero/Tare Errors

Zero/Tare Errors

Zero/Tare Errors

Zero/Tare Errors

2.0

Balance is unable to stabilize within time limit after zero/taring. Environment is
too hostile or balance needs recalibration.

Calibration Errors

Calibration Errors

Calibration Errors

Calibration Errors

Calibration Errors

3.0

Incorrect or no calibration weight used for calibration. Recalibrate with correct
weights.

RS232 Errors

RS232 Errors

RS232 Errors

RS232 Errors

RS232 Errors

4.0

Bad RS232 frame. Check RS232 menu parameters and correct.

4.4

RS232 buffer is full (if installed). May occur if no printer or computer is
connected to the interface. To clear buffer, turn balance off or enter Print menu
and select END.

4.5

Function is disabled by the Lock switch.

User Errors

User Errors

User Errors

User Errors

User Errors

7.2

Number outside of display capacity.

Over-Under Load Errors

Over-Under Load Errors

Over-Under Load Errors

Over-Under Load Errors

Over-Under Load Errors

8.0

Hardware error causing an internal weight signal which is too low. Check if pan
or pan support is off. If not, the balance must be serviced.

8.1

Hardware error caused by an internal weight signal which is too high. Check
load on the platform which may be excessive. If error persists, the balance must
be serviced.

8.2

Power-on load out of specification: Balance was turned on with load on pan or
pan off balance. No load may be on pan when turned on and pan must be in
place.

8.3

Rated capacity exceeded. Remove excessive weight from pan.

8.4

Underload condition on balance. Check that the proper pan and pan support are
installed.