Linearity calibration – Ohaus AS Series Electronic Balances User Manual

Page 49

background image

15

Linearity Calibration

Linearity Calibration

Linearity Calibration

Linearity Calibration

Linearity Calibration

1.

Press and hold

until CAL is

displayed, then release it. Balance

will display SPAN.

2.

Press

and the display will show

LIN.

3.

Press

to start the Linearity

Calibration Procedure.

4.

When

is released, C 0g will be

displayed, indicating that no weight

should be on the pan.

5.

Press

. The display will show

-C- momentarily, followed by the val-

ue of the weight which must be placed

on the pan. Do not disturb the bal-

ance when -C- is displayed. Distur-

bances will result in improper calibra-

tion.

6.

Place the required weight on the pan.