12 PREPARATION

FOR

INSTALLATION

 Select the optimal sampling point in accordance with the generally applicable guidelines or

consult the competent persons.

 Locate the sampling point in such a way that there is adequate space for inserting and

removing the probe and pay attention to the insertion length of the probe tube.

 Make certain that the probe is easily accessible so that you can carry out any subsequent

maintenance work without trouble.

 Locate the probe connections in such a way that the connections' temperature is always above

the acid dew point in order to avoid corrosion and blockage problems. If this is not possible, a
heated SP35/SP30 probe tube is recommended for cold connections.

 If the ambient temperature in the area of the connections is >80°C as a result of radiated heat,

a radiated-heat deflector must be mounted to protect the probe.

 The connection's mounting flange connection should comply with DN65 PN6. If other

connection sizes are required, a special adapter flange /S010 can be supplied as an option.

 Before mounting, the probe must be adapted to the existing operating conditions.


The existing operational parameters are to be checked accordingly prior to commencing mounting
work:

weatherproof mounting position

_____present _____set

Under / over pressure situation

mbar bar

Process temperature

°C, Min. °C, Max.

Dust loading

g/m³

Dust composition - grain size

µm

Gas composition

corrosive toxic explosive

Which parameters should be
measured, e.g. O

2

, CO, SO

2

, NO

X

,..,

Vol.% mg/Nm³ ppm

Required amount of gas

l/hr,
Min.

l/hr,
Max.

Necessary T

90

time

sec.



13 MOUNTING

M&C

SP180-H probes are designed for stationary use and if properly selected and mounted a long

service life and minimum maintenance are guaranteed. It is advisable to mount the probe in an
operational position which is at a 10% inclination to the process. See also Figure 2 for the following
description.

 Screw either the probe tube directly into the ¾" inner thread  in the flange of the probe with

the ¾" flat gasket  and tighten.

 For mounting of the SP180-H at the sampling flange loosen the metal clamp  around the heat

insulating protection cover and remove the cover .

 If the probe connection does not correspond to the standard flange connection DN65 PN6, an

optionally supplied adapter flange  should be mounted to the probe (see Figure 2).

 If the heated probe tube type SP30/35 is used then the probe is to be screwed to its flange

(with welded threaded bolt). First insert the flange seal  between the two flanges.

 Attach the flange seal  to the probe connection.
 Insert mounted probe tube into the connection piece and fasten the probe using the screws

and nuts supplied.