TECHNICAL DATA
8.1
FOR BASIC COOLER EC-L WITHOUT HEAT EXCHANGER
Gas cooler series EC
version EC-L
Part No. for basis cooler without
heat exchanger, 230V 50Hz
02K4000X (a)**
Part No. for basis cooler without
heat exchanger, 115V 50-60Hz
prepared for 1 stream
Number of heat exchangers
2
Ambient temperature
+10 up to +45 °C
Sample outlet dew point
range of adjustment: +2 °C ..... +7 °C, factory setting: +5 °C
Dew point stability
at constant conditions < ±0,1 °C
Sample inlet temperature
*** max. 180 °C
Sample inlet dew point
*** max. 80 °C
Total cooling capacity
144 KJ/h at +10 up +45 °C ambient
Ready for operation
< 15 min
Main power connection / power
consumption
230V AC* or 115V AC**(a) -15%/+10%, 50/60 Hz, max. 200VA
start up current: 230V 50Hz = 2,5A / 115V 60Hz = 4,5A
Electrical connection
terminals 2,5mm
2
, 2x PG11 cable gland (with FM-approval conduit-hub
1/2” NPT)
Status alarm
2 contacts, potential free contact rating 250VAC, 2A, 500VA, 50W,
alarm point <+2 °C and >+8 °C*
Case protection / Electrical standard IP20 EN60529 / EN 61010
Method of mounting / Case colour
wall mounting / case colour grey, RAL 9003
Dimension / Weight
210x311,5x300 mm (w x h x d) / 17,6 kg (with 2 SR25.2 and 2 glass
heat exchangers)
8.2
OPTIONS FOR BASIC COOLER EC-L
Options for basic cooler
EC-L
Heat exchanger type
ECL-G
ECL-PV
ECL-SS
Part No
97K0605
97K0610
97K0600
Material of heat exchanger
Duran glass
PVDF
st. steel 316Ti
Max. gas flow rate per heat
exchanger
250Nl/h
250Nl/h
250Nl/h
Gas press. max. bar abs.
3)
2 / 3
2)
3
10*
Sample gas connection
GL18
for Ø6mm a.d. tube*
Tube ø 6 mm*
Tube ø 6 mm
Condensate connection
GL25 for Ø12mm tube*
Ø8mm or Ø10mm**
G3/8’’i
G3/8’’i or
3/8 NPT**
P at max. flow rate
1mbar
1mbar
1mbar
Stagnant space approx.
50ml
ml
ml
Peristaltic pump SR25.2
1 pc. incorporated in the cooler, compl. installed, Part No.: 01P9125 cooler
weight plus 0,6 kg per pump
* Standard, other version on request.
** Option
***
Maximum values in technical data’s must be rated in consideration of total cooling capacity at 25 °C ambient temperature..
2) With GL-Connecting adapter.
3) With SR25.2 max. 2 bar abs.