beautypg.com

Technical data, Applications, Description – M&C TechGroup SP2100-H Operator's manual User Manual

Page 8

background image

8

Gas sampling and gas conditioning technology

2-1.0-ME

7

TECHNICAL DATA

Probe series SP®

compact version SP 2100-H

Part No.

10S1000

Sample tube

Type SS, stainless steel 316Ti, length 1 m*

Sample temperature

max. 600 °C*

Sample pressure

0.4 to 2 bar abs.

Ambient temperature

-20 °C to +60 °C

Recommended for dust loadings

up to 2 g/m

3

*

Filter chamber volume

120 ml

Filter element

S-2K 150, filter porosity 2

μm, ceramic

Probe heating temperature

+180 °C self-regulating

Ready for operation

after 2 hours

Temperature alarm contact, alarm point

<160 °C, NO

Temperature alarm, contact rating

250V-3A AC, 30V-3A DC

Connection sample outlet

1/4"-NPT inside, for max. 10 mm tube connectors

Power supply

110 up to 240V 50 / 60Hz

Power consumption

start up: 400VA, usual: 100VA, (fuse 6A)

Electrical connection

terminals max. 2,5 mm

2

, 2x PG11 cable gland

Electrical equipment standard

EN 61010, EN 60335-1

Degree of protection

IP54, EN 60529

Mounting flange

DN65 PN6,B stainless steel 316Ti

Material of sample contacting parts

SS 316 / 316Ti, FPM, ceramic

Weight

9 kg

Options

not available

P und T90 at flow of:

100

200

500

1000

Nl/h

P pressure loss with new filter element S-2K150:

0,007

0,011

0,020

0,035

bar

T90 time-without sample tube/prefilter-

6,0

3,5

1,0

<0,5

sec



8

APPLICATIONS

The patented

M&C SP2100-H

gas sample probe is used for continuous gas sampling in processes

with dust densities of up to 2g/m³, operating pressure up to max. 2 bar abs., temperatures of up to a
maximum of 600°C or higher gas humidity. Thanks to its compact design it requires only limited space.
It has to be mounted in a weatherproof position.


9

DESCRIPTION

The sample probes are designed for easy installation, reliable operation and trouble-free maintenance.
Filter elements can be changed without the need for tools and without disconnecting the sample line,
the filter chamber can be cleaned easily and the probe tube can be cleaned without removing the
probe:
these are just a few of the many advantages offered by this probe.

The 2 µm ceramic filter element, which has a large surface, is located in a stainless steel filter
housing.

The gas sample probe is heated up to + 180°C using a special self-regulating heating element. No
thermostat or temperature limiter is necessary.

A separate thermo switch is provided for under-temperature monitoring (< 160°C,

NO

).