ELECTRICAL CONNECTION
W A R N I N G !
When connecting the equipment, please ensure that the supply
voltage is identical with the information provided on the model type
plate.
N O T E !
Attention must be paid to the requirements of IEC 364 (DIN VDE
0100) when setting high-power electrical units with nominal
voltages of up to 1000 V, together with the associated standards
and stipulations.
In any case we recommend the use of temperature resistant cable !
A main switch and matching fuse must be provided externally!
The main circuit must be equipped with a fuse corresponding to
the nominal current (over current protection); for electrical details
see technical data.
It is recommended to use the low temperature alarm. In case of an
alarm the flow can be stopped and the components downstream
the probe are safe for demage.
Remove the lid of the connection socket. The electrical connection layout is located in the lid.
Insert the mains cable (min. 3 x 1.5 mm²) through the threaded cable connection and connect
to the appropriate terminals.
Insert the signal cable through the threaded cable connection and connect to the appropriate
terminals. Screw lid back on.