APPLICATIONS

In analysis technique, often temperatures must be kept above the sample gas dew point. Therefore, it
is absolutely necessary to avoid cold spots. In order to ensure this, the temperature regulated M&C
three-way valves V3/2-

H… and MV3/2 H are used for the cut-off or switching-over of sample gases

and test gases up to an operating temperature of 180 °C.


9

DESCRIPTION

The M&C electrically heated valves are fixed on a mounting plate, decoupled from heat and covered
with an insulated enclosure. The heater consists of a heating element with high capacity. The tem-
perature is adjustable on the integrated thermostat up to 180 °C with high temperature limiter and low
temperature alarm. The heat insulated enclosure with a bushing for the dial thermometer is equipped
with quick acting bentlever closures. The connection box with integrated thermostat is installed outside
the enclosure on the mounting plate. Version V3/2-H/PE has got an additional connecting box for the
contact output of the position identification. The power supply for the solenoid valves version MV3/2-H
is 24V DC and the additional connecting box is installed on the mounting plate. In order to prevent
cold spots, the connecting fittings are heated by means of thermal conducting jaws. The electrically
heated sample lines type 3/4/5-N/M/H are fixed with mounting brackets.


10

RECEIPT OF GOODS AND STORAGE

Please take the heated valve and possible special accessories carefully out of the packaging

material immediately after arrival, and compare the goods with the items listed on the delivery
note!

Check the goods for any damage caused during delivery and, if necessary, notify your trans-

port insurance company without delay of any damage discovered.

N O T E !


The equipment should be stored in a protected, frost-free room!



11

PREPARATION AND INSTALLATION

Locate the heated 3-way-valves type V3/2-

H… and MV3/2-H in such a way that there is adequate

space for removing the cover and connecting the sample lines. Fix the aluminium plate with 4 screws.
Make certain that the heated 3-way-valves are easily accessible so that you can carry out any subse-
quent maintenance work without trouble.