2 electrical connections – M&C TechGroup SR25.1_Ex Operator's manual User Manual
Page 12

12
Gas sampling and gas conditioning technology
3-7.1-ME
N O T E !
The tightness of the connections can only be guaranteed if the
connection hose has a straight rim (hose cutter).
C A R E !
Aggressive condensate is possible!
Wear protective glasses and proper protective clothing!
11.2 ELECTRICAL
CONNECTIONS
Connect the cables for the power supply as follows:
Figure 5 Connection of the distribution voltage
D A N G E R !
Incorrect system voltage can damage the unit. When establishing
connections, check that the system voltage corresponds with the
voltage shown on the type plate!
The supply voltage is only allowed to deviate max. +6% resp. -10%
from the indication on the model type plate.
N O T E !
Attention must be paid to the requirements of IEC364 (DIN VDE 0100)
when setting high-power electrical units with nominal voltages of up
to 1000V, together with the associated standards and stipulations.
The electrical connection of the SR25.1/Ex must be done in a separate EExe box with the cable gland
M12x1 included:
230 V - blue and red
115 V - blue and white.
Power
Power