Laser:cal:mdi, Laser:cond – Lightwave Communications LDX-3232 User Manual
Page 62

C O M M A N D R E F E R E N C E
LDX-3232 Device-Dependent Command Reference
46
LDX-3232
C H A P T E R
4
The LASer:CAL:MDI? query is used to determine that the LDX-3232 is ready for a value
to be entered during the calibration cycle of the LASer:CAL:MDI mode.
Syntax Diagram
Notes
After this query is issued and a response of 1 is received, the LDX-3232 be ready for the
user to enter a photodiode current value via the LASer:MDI command (see Chapter 6).
Examples
"LAS:CAL:MDI?" -response: 1, means the LDX-3232 is ready for the user to enter a
photodiode current value via the LASer:MDI command.
The LASer:COND? query returns the value of the status condition register of the LASER
operations.
Syntax Diagram
-where the response is the sum of the following:
Notes
The LASER conditions that are reported to the status byte are set via the
LASer:ENABle:COND command.
The Open circuit condition is only present while the LASER output is on, and when the
hardware detects this condition, it will turn the LASER output off. Therefore, the Open
LASer:CAL:MDI?
F
RONT
P
ANEL
R
EMOTE
LASer:COND?
F
RONT
P
ANEL
R
EMOTE
1
LASER Current Limit
256
Output is shorted
2
LASER Voltage Limit
512
Output is outside tolerance limit
4
N/A
1024
Output on/off state
8
Power limit
2048
Ready for calibration data state
16
Interlock disabled
4096
N/A
32
N/A
8192
N/A
64
Compliance Voltage Limit
16384
N/A
128
Open circuit
32768
N/A
:
LAS
:
CAL
IPD
:
?
MDI
:
LAS
:
COND
?