Controlled power-
down sequence

Before performing the controlled power-down sequence, determine if
the error can be addressed from the host. If so, perform the required
host recovery procedures. If the error cannot be addressed from the
host, or if the scanner’s error message cannot be cleared when the
host recovery procedures are performed, continue with the following
power-down sequence.

1. Wait for all information currently held in the buffer to be transferred to

the host system.

2. If necessary, prepare the host and corresponding system

components for a scanner power-down, per host system
instructions.

3. Turn off the scanner power for at least 5 seconds, then turn the

power back on again.

4. If necessary, determine the last image transferred to the host, reset

the image address and image level to the correct values, and rescan
all documents whose images have not been transferred to the host.

5. If required, calibrate the scanner before rescanning.

6. If the error condition is not cleared using the controlled power-down

sequence, contact Service.