beautypg.com

Selftest and calibration overview, Purpose of calibration, When to calibrate – HP 2040D User Manual

Page 9

background image

Series 2040 Test System

Series 2040 Maintenance Manual V2.00

Selftest & Calibration

8

SELFTEST AND CALIBRATION OVERVIEW

Calibration on the Series 2040 Functional Test System is completely

automated. All calibration is based on high precision references built into the

system Testhead and the Selftest Assembly. A high precision Temperature

Compensated Crystal Oscillator (TCXO) on the Time Measurement System

Board (TMS) is used as a frequency/time reference. A precision digital to

analog converter (TDAC) contained within the Selftest Assembly is used as a

voltage reference for system calibration.

Purpose Of Calibration

Calibration of the functional tester is necessary to obtain gain and offset terms

for all signal paths that exist within the system, and to determine timing for the

MDE’s (Measurement Display Electronics) sweeps, delays, triggers, and

measurement mark. Since potentiometers are used in only a few instances

where factory calibration is required, the majority of the calibration is done by

taking measurements and calculating calibration terms which can be

referenced at a later time. These terms are used by the hardware functional

calls to account for gains, offsets, and timing discrepancies, and their use is

transparent to the user.

In all calibration programs, limits are set on what values any given calibration

term may take on. If a calculated value is outside of the allowable range, a

failure will result and be displayed/printed to the designated output device .

When To Calibrate

Selftest is the best guide to determine if calibration is required. The battery of

Selftest programs is designed to test all functions of the 2040 to their

specifications. If any failures occur during Selftest (especially borderline

failures), calibration should be carried out before proceeding to troubleshoot

the system.

Running through Selftest periodically can tell you if calibration is or is not

required. If you find that all Selftest programs run without failures, there is no

need to recalibrate any portion of the tester. The recommended guideline is to

calibrate every 3 months.

Whenever a temperature change of more than five (5) degrees Celsius occurs

or a Testhead board swap is made, the 2040 should be recalibrated.