Series 2040 Test System

Series 2040 Maintenance Manual V2.00

Diagnostics

31

TESTHEAD - PATCHBOARD - UUT P/S BUILT-IN DIAGNOSTICS

When a malfunction is suspected in either the Testhead Internal or

Patchboard Power Supply system, the LED status indicators should be

examined. The top LEDs are labeled for their respective power supplies as

shown on page 14. The legend for the four lower LEDs is shown below. In

case of a power supply malfunction, the power supply that has shut down first

will be indicated by its respective LED being extinguished.

If you are diagnosing the case of no power at the Patchboard, and the status

indicators, the lower four LEDs, are all lit, then an over/under voltage fault has

occurred in a Patchboard power supply. The power supply causing the fault

will have it’s respective LED extinguished. Power could also have been

removed from the Patchboard without generating a fault condition. In this

case, the reason that the power has been removed is indicated by one or two

of the lower four status LEDs being extinguished.

CASE

PROX

PBON

Ext.Fault

Reset

Power Call

on

off

on

on

TCLEAR

on

off

on

off

Patchboard removed

off

on

on

on

UUT Power Fault

on

on

off

on

When a power supply cannot be restored by resetting the unit, the circuit

breaker for the offending supply should be checked for continuity. A graphic of

the circuit breakers on the edge of the Testhead Power Supply Controller

board is shown on page 14 of this section. If any of these breakers continually

trips, the associated power supply must be replaced.