Calibration – HP 2040D User Manual

Page 17

background image

Series 2040 Test System

Series 2040 Maintenance Manual V2.00

Selftest & Calibration

16

CALIBRATION

Whenever a Testhead board is

replaced, or any time that Selftest

determines a failure, a calibration

must be performed. If a unit

passes the complete Functional

Selftest routine, a calibration is

not necessary. When Calibrate is

selected from the Selftest

Executive, the screen shown to

the right is displayed.

The Calibrate menu operates

under the same basic rules as the

Functional menu. For installation

purposes, use the mouse to select the “All Tests” box, and select the “1"

button for the number of times to be run. (See page #6). Ensure that the

Printer and Fail Data are selected, so that any failures will be printed

automatically.

The following is a list of the calibration programs located in the Calibrate

menu in the Selftest Executive, with the three most likely boards or assemblies

that could cause failures:

Test Name

#1

#2

#3

#1 AMS_RMUX_c AMS

RMUX

MSP/ISO

#2 AMS_SIG3_c

AMS

MSP/ISO

RMUX

#3 AMS_DIFF_c

AMS

RMUX

ICAM

#4 DMS_c

DMS

#5 ARB_v_c

ASB

MSP/ISO

AMS

#6 DA_v_c

ASB

MSP/ISO

AMS

#8 DIO_c

DIO TH CONT/CPU

Selftest

#9 ADIO_c

ADIO TH CONT/CPU

Selftest

#10 ADIO_dac_c

ADIO

AMS

RMUX

#12 MDE_c

MDE

TMS

AMS

#13 MDE_TGDC_c MDE

TMS

AMS

#15 POWER_c

UUT Cont

AMS/RMUX Selftest

#17 ISOAmp_c

MSP/ISO

AMS

RMUX