Series 2040 Test System

Series 2040 Maintenance Manual V2.00

Certification

47

When this portion of the certification is complete, the TCXO frequency must

be checked. The procedure for the frequency certification is identical, except

for the final screen displaying the frequency counter. The user should wait

until the frequency is stable, and then click on certify to start the test. Again,

the results and a pass/fail condition are displayed.

For the portion dealing with the TDAC, the measured value must not vary

from the programmed value by more than the designated limits or a failure

will result. If this happens, a manual adjustment must be made. Contact

Digalog Systems should this occur.