Magnetic valve, Removing the magnetic valve diaphragm and spring – HP 403349-003 User Manual
Page 118
Technician repair removal and replacement procedures 118
9.
Replace the condensation pump and sensors ("
Replacing the condensation pump and sensors
10.
Relocate the power inlet box back into position ("
Relocating the power inlet box back into position
11.
Replace the bottom fan unit ("
12.
Restore water flow to the MCS unit ("
13.
Complete the operation checklist (on page
Magnetic valve
The magnetic valve is a component of the water group located at the bottom interior of the MCS unit.
The following tools are required for installation:
•
T-25 Torx driver
•
Phillips screwdriver
•
8-mm Allen wrench
Removing the magnetic valve diaphragm and spring
1.
Power down the MCS unit ("
2.
Disconnect the main water lines at the quick disconnect couplings.
3.
Remove the bottom fan unit ("
4.
Relocate the power inlet box ("
Relocating the power inlet box
5.
Remove the air deflector plate ("