User’s Guide for DS1000B series

2-50


To Set up the Sampling System


Figure 2-59 shows, the Acquire button at the MENU of the front panel.

Figure 2-59

The Acquire setup button

Press Acquire button, the interface menu shows as follows:

Figure 2-60 Table 2-37 The Acquire menu

Menu

Settings

Comments

Acquisition

Normal
Average
Peak Detect

Normal Acquisition mode.
Average Acquisition mode.
Peak Detect Acquisition mode.

Averages

[1]

2 to 256

Set average times by N-th power
of 2 from 2 to 256.

Sampling

Real-Time
Equ-Time

Real-time sampling mode.
Equivalent sampling mode.

Sinx/x

[2]

ON
OFF

Set the interpolation as Sinx/x.
Set the interpolation as linearity.

Sa Rate

[3]

Display sampling rate.

NOTES:

[1] This function appears when acquisition as “Average”.

[2] This function appears when sampling as “Real-Time”.

[3] Sample rate channels will be displayed respectively in alternative trigger mode.

Acquire Setup Button