beautypg.com

RIGOL DS1000B User Manual

Page 158

background image

RIGOL

User’s Guide for DS1000B Series

5-4

Trigger
Trigger Sensitivity

0.1div-1.0div (adjustable)

Trigger Level Range

Internal ±6 divisions from center of screen
EXT

±1.2V

EXT/5

±6V

Trigger Level Accuracy
(typical) applicable for
the signal of rising and
falling time ≥20ns

Internal

±(0.3div × V/div)(±4 divisions from
center of screen)

EXT

±(6% of setting + 40 mV)

EXT/5

±(6% of setting + 200 mV)

Trigger Offset

Normal mode: pre-trigger(storage depth/(2×sampling)
rate), delayed trigger 1s
Slow Scan mode: pre-trigger 6div, delayed trigger 6div

Trigger Holdoff range

100ns~1.5s

HF Rejection

100kHz ±20%

LF Rejection

10kHz ±20%

Set Level to 50%
(typical)

Input signal frequency ≥50Hz

Edge Trigger
Edge trigger slope

Rising, Falling, Rising + Falling

Pulse Trigger
Trigger condition

(>, <, =) Positive pulse, (>, <, =) negative pulse

Pulse Width range

20ns ~10s

Video Trigger
Video standard & line
frequency

Support standard NTSC, PAL and SECAM broadcast
systems. Line number range: 1~525 (NTSC) and 1~625
(PAL/SECAM)

Pattern Trigger
Pattern setup

H, L, X,

,

Alternate Trigger
Trigger on CH1, CH2,
CH3, CH4

Edge, Pulse, Video


Measurements

Cursor

Manual Voltage difference between cursors (∆V)

Time difference between cursors (∆T)