RIGOL DS1000B User Manual
Page 42
RIGOL
User’s Guide for DS1000B series
2-12
5. To change the Volts/Div settings
The Volts/Div control has Coarse or Fine configuration. The Vertical
Sensitivity is 2mv/div - 10V/div.
Coarse: It is the default setting of Volts/Div in a 1-2-5-step sequence from
2mV/div, 5mV/div, 10mV/div, to 10 V/div.
Fine: This setting changes the vertical scale to small steps between the coarse
settings. It will be helpful to adjust the waveform in smooth steps.
Figure 2-12 Fine Adjustment
Coarse/Fine Shortcut key
To change Coarse/Fine setting, not only by menu but also by pressing
vertical
knob.
Fine Adjustment Data
Fine
Adjustment
See also other documents in the category RIGOL Measuring instruments:
- DS1000E (2 pages)
- DS1000E (2 pages)
- DS1000E (2 pages)
- MSO/DS4000 Series (2 pages)
- MSO/DS4000 Series (63 pages)
- MSO/DS4000 Series (45 pages)
- DS6000 Series (79 pages)
- MSO/DS4000 Series (42 pages)
- MSO/DS1000Z Series (2 pages)
- MSO/DS4000 Series (44 pages)
- MSO/DS1000Z Series (12 pages)
- MSO/DS4000 Series (66 pages)
- MSO/DS4000 Series (49 pages)
- DS6000 Series (280 pages)
- DS1000B (9 pages)
- DS1000B (152 pages)
- DS1000E (36 pages)
- DS1000E (10 pages)
- DS1000E (137 pages)
- DS1000E (166 pages)
- MSO/DS4000 Series (16 pages)
- MSO/DS4000 Series (68 pages)
- MSO/DS2000A Series (2 pages)
- MSO/DS2000A Series (77 pages)
- MSO/DS2000A Series (317 pages)
- MSO/DS1000Z Series (72 pages)
- MSO/DS1000Z Series (8 pages)
- MSO/DS1000Z Series (270 pages)
- M300 (66 pages)
- M300 (169 pages)
- DM3058E (138 pages)
- DM3058E (47 pages)
- M300 (10 pages)
- DM3058E (192 pages)
- M300 (89 pages)
- M300 (144 pages)
- DSA1030 (12 pages)
- DSA1030 (2 pages)
- DSA1030 (18 pages)
- DSA1030 (2 pages)
- DSA1030 (1 page)
- DSA1030 (2 pages)
- DSA1030 (5 pages)
- DSA1030 (67 pages)