User’s Guide for DS1000B series

2-78


1. Mask setting

Press Utility→Pass/Fail→Mask Setting to go to the following menu.


Figure 2-103 Table 2-57 The Mask setting menu (Page 1/2)

Menu

Settings Comments

X Mask

Set horizontal clearance to the waveform
(0.04div-4.00div).

Y Mask

Set vertical clearance to the waveform
(0.04div-4.00div).

Create
Mask


Create a test mask according to the above
clearance.

Location

Internal
External

Set the memory location of the mask
files.



Figure 2-104 Table 2-58 The Mask setting menu (Internal, Page 2/2)

Menu

Settings

Comments

Save

Store created test mask into internal
memory.

Recall

Recall mask setting file from internal
memory.

Imp./Exp.

Go to import/export menu (same as
REF import/export menu. see table
2-15).