Technical specifications, Technical specifications -2 – RIGOL DS1000B User Manual
Page 156
RIGOL
User’s Guide for DS1000B Series
5-2
Technical Specifications
Acquisition
Sampling Modes
Real-Time
Equivalent-time
Sampling Rate
2 GSa/s (half channel
[1]
)
1 GSa/s (each channel)
50 GSa/s
[2]
Averages
N time acquisitions, all channels simultaneously, N is
selectable from 2, 4, 8, 16, 32, 64, 128 and 256
Inputs
Input Coupling
DC, AC, GND
Input Impedance
1MΩ±2.0%
The input capacity is 18pF±3pF
Probe Attenuation
Factors
0.001X, 0.01X, 0.1X, 1X, 2X, 5X, 10X, 20X, 50X, 100X,
200X, 500X, 1000X
Maximum Input Voltage
Maximum Input Voltage of the analog channel:
CAT I 300Vrms, 1000Vpk; transient overvoltage
1000Vpk
CAT II 100Vrms, 1000Vpk
RP2200 10:1, CAT II 300Vrms
RP3300A 10:1, CAT II 300Vrms
Time delay between
channel (typical)
500ps
Horizontal
Sample Rate Range
3.65Sa/s-2GSa/s (Real-Time),
3.65Sa/s-50GSa/s (Equivalent-time)
Waveform interpolation Sin(x)/x
Memory Depth
16k samples when horizontal timebase is 20ns/div or
lower and 8k samples when horizontal timebase is
50ns/div or higher for half channel
[1]
8k samples for each channel
Scan speed Range
(Sec/div)
1ns/div~50s/div, DS1204B
2ns/div~50s/div, DS1104B
5ns/div~50s/div, DS1074B
1-2-5 Sequence