Waveform:points:mode – RIGOL DS1000B User Manual
Page 84
RIGOL
Programming Guide for DS1000B Series
2-72
For details about storage format of waveform points, please refer to Page 2-76:
Peak Detect
4. :WAVeform:POINts:MODE
Command Format:
:WAVeform:POINts:MODE
:WAVeform:POINts:MODE?
Function:
This command sets the mode of waveform points.
NORMal, MAXimum or RAW.
NOTE: What will be returned by :WAVeform:POINts? in different modes:
NORMal: Return data points currently display on the screen (600 points).
RAW: Return the data points of the memory data (in STOP state). In RUN state, no
data are returned, the system error code is 67 which indicate system condition is not
met and the execution failed.
MAXimum: Return the maximum valid data points in current state. In RUN state, the
screen data points are returned, while in STOP state, the memory data points are
returned.
NORmal
RAW
MAX
Normal
/Average
Peak
Detect
In RUN state,
MAX is the same
with NORMal;
in STOP state,
MAX is the same
with RAW.
MATH
600
1200
600
FFT
500
500
500
CHx
600
1200
8192
Half-Channel
CHx
600
1200
16384
Returned Format:
The query returns NORMal, MAXimum or RAW.
Example:
:WAV:POIN:MODE NORM Set the mode as NORMal.
:WAV:POIN:MODE? Return NORMal.