beautypg.com

Waveform:points:mode – RIGOL DS1000B User Manual

Page 84

background image

RIGOL

Command Systems

Programming Guide for DS1000B Series

2-72

For details about storage format of waveform points, please refer to Page 2-76:
Peak Detect


4. :WAVeform:POINts:MODE

Command Format:
:WAVeform:POINts:MODE
:WAVeform:POINts:MODE?

Function:
This command sets the mode of waveform points. can be:
NORMal, MAXimum or RAW.

NOTE: What will be returned by :WAVeform:POINts? in different modes:

NORMal: Return data points currently display on the screen (600 points).

RAW: Return the data points of the memory data (in STOP state). In RUN state, no

data are returned, the system error code is 67 which indicate system condition is not
met and the execution failed.

MAXimum: Return the maximum valid data points in current state. In RUN state, the

screen data points are returned, while in STOP state, the memory data points are
returned.

NORmal

RAW

MAX

Normal
/Average

Peak
Detect

In RUN state,
MAX is the same
with NORMal;
in STOP state,
MAX is the same
with RAW.

MATH

600

1200

600

FFT

500

500

500

CHx

600

1200

8192

Half-Channel
CHx

600

1200

16384

Returned Format:
The query returns NORMal, MAXimum or RAW.

Example:

:WAV:POIN:MODE NORM Set the mode as NORMal.
:WAV:POIN:MODE? Return NORMal.