Command Systems

Programming Guide for DS1000B Series

2-30

NORMal: Acquire waveform when trigger occurred;

SINGle: Execute once trigger when all the condition are marched and stop.


Returned Format:
The query returns AUTO or NORMAL, SINGLE.

Example:
:TRIG:EDGE :SWE AUTO Set the trigger type as AUTO.
:TRIG:EDGE :SWE?

Return AUTO.


5. :TRIGger:SENSitivity

Command Format:
:TRIGger:SENSitivity
:TRIGger:SENSitivity?

Function:
The command sets the trigger sensitivity. The range is 0.1div~1div.

Returned Format:
The query returns the value of trigger sensitivity, and the unit is div.

Example:
:TRIG:SENS 0.2 Set the trigger sensitivity as 0.2div.
:TRIG:SENS? Return 2.000e-001.

6. :TRIGger:COUPling

Command Format:
:TRIGger:COUPling {DC|AC|LF}
:TRIGger:COUPling?

Function:
The command sets the coupling mode.
 DC: Allow all signals pass;
 AC: Reject DC signals and attenuate the signal below 10Hz;
 LF: Reject DC signals and attenuate the signals below 8kHz.