Command Systems

RIGOL

Programming Guide for DS1000B Series

2-53

Function:
The command sets the vertical offset. The may be 1, 2, 3 or 4.
Scale≥250mV, : -40V~ +40V;
Scale<250mV, : -2V ~ +2V.

Returned Format:
The query returns the value of offset, and the unit is V.

Example:
:CHAN2:OFFS 20 Set the vertical offset of channel 2 as 20V.
:CHAN2:OFFS? Return 2.000e001.


6. :CHANnel:PROBe

Command Format:
:CHANnel:PROBe
:CHANnel:PROBe?

Function:
The command sets the attenuation factor of probe. The may be 1, 2, 3 or 4,
and the may be 0.001 X, 0.01 X, 0.1 X, 1X, 2 X, 5X, 10X, 20 X, 50X, 100X,
200 X, 500X or 1000X.

Returned Format:
The query returns the value of attenuation factor.

Example:
:CHAN2:PROB 10X Set the annenuation factor of channel 2 as 10X.
:CHAN2:PROB?

Return 10X.



7. :CHANnel:SCALe

Command Format:
:CHANnel:SCALe
:CHANnel:SCALe?

Function: