RIGOL

Command Systems

Programming Guide for DS1000B Series

2-40

3. :TRIGger:ALTernation:TYPE

Command Format:
:TRIGger:ALTernation:TYPE [,]
:TRIGger:ALTernation:TYPE? []

Function:
The command sets the trigger type. The may be EDGE, PULSe or VIDEO,
and the may be SOURceA or SOURceB, and the source A and B are
varying with the current alternation channel.

Returned Format:
The query returns EDGE or PULSE, VIDEO.


Example:
:TRIG:ALT:TYPE EDGE,SOURB Set the trigger type as edge trigger.
:TRIG:ALT:TYPE? SOURB

Return EDGE.

4. :TRIGger:ALTernation:TimeSCALe

Command Format:
:TRIGger:ALTernation:TimeSCALe [,]
:TRIGger:ALTernation:TimeSCALe? []

Function:
The command sets the time scale of current channel. The range is
2ns~20ms, and the may be SOURceA or SOURceB, and the source A
and B are varying with the current alternation channel.
In NORMAL mode, different types of instruments have different sweep ranges:

DS1204B, range:1ns/div~50s/div.
DS1104B, range:2ns/div~50s/div.
DS1074B, range:5ns/div~50s/div.

Returned Format:
The query returns the value of time scale, and the unit is s.

Example:
:TRIG:ALT:TSCAL 0.001,SOURB Set the time scale as 1ms.