beautypg.com

Trigger:video:standard, Trigger:video:line – RIGOL DS1000B User Manual

Page 49

background image

Command Systems

RIGOL

Programming Guide for DS1000B Series

2-37

3. :TRIGger:VIDEO:STANdard

Command Format:
:TRIGger:VIDEO:STANdard {NTSC|PALSecam}
:TRIGger:VIDEO:STANdard?

Function:
The command sets the video standard as NTSC or PAL/SECAM.

Returned Format:
The query returns NTSC or PAL/SECAM.

Example:
:TRIG:VIDEO:STAN PALS Set the video standard as PAL/SECAM.
:TRIG:VIDEO:STAN?

Return PAL/SECAM.


4. :TRIGger:VIDEO:LINE

Command Format:
:TRIGger:VIDEO:LINE
:TRIGger:VIDEO:LINE?

Function:
The command sets the number of sync specified line. In NTSC standard, the
range is 1~525; in PAL standard, the range is 1~625.

Returned Format:
The query returns the number of specified line.

Example:
:TRIG:VIDEO:LINE 25 Set the number of sync specified line as 25.
:TRIG:VIDEO:LINE?

Return 25.