RIGOL

Command Systems

Programming Guide for DS1000B Series

2-58

1. :MEASure:CLEar

Command Format:
:MEASure:CLEar

Function:
The command clears the current measurement parameters.



2. :MEASure:VPP?

Command Format:
:MEASure:VPP? []

Function:
The command measures the Peak-Peak value of signal from . The
may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.

NOTE: “<>” in indicates the source is the parameter that must be set in the
commond; and “[ ]” indicates the parameter can be set or not set according to your demand.
The followings are the same, don’t repeat. Please refer to chapter 1 Symbol Description
about the related explanations.

Returned Format:
The query returns as 5.280e000, and the unit is V.



3. :MEASure:VMAX?

Command Format:
:MEASure:VMAX? []

Function:
The command measures the maximum of signal from . The
may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.

Returned Format:
The query returns as 2.640e000, and the unit is V.