beautypg.com

Timebase:delayed:offset, Timebase[:main]:scale – RIGOL DS1000B User Manual

Page 35

background image

Command Systems

RIGOL

Programming Guide for DS1000B Series

2-23

:TIM:OFFS?

Return 1.000e000.



3. :TIMebase:DELayed:OFFSet

Command Format:
:TIMebase:DELayed:OFFSet
:TIMebase:DELayed:OFFSet?

Function:
The command sets the timebase offset of delayed scan, that is the offset of the
waveform position relative to center screen.
In NORMAL mode, : 1s ~ memory capacitance;
In STOP mode, : -500s ~ +500s;
In SCAN mode, :
( -6*MainScale + 6*DelayedScale

) ~(6*MainScale – 6*DelayedScale)

Scale is the current horizontal scale, and the unit is s/div.

NOTE①: In Delayed mode, only Delayed offset can be changed but for Main offset.
Thereinto:
the time range: +/- 6*MainScale;
The length of time is 12*DelayedScale;
So Delayed Offset range is:
(-6*MainScale + 6*DelayedScale) ~ (6*MainScale-6*DelayedScale).
For example: When Main 5ms, Delay 2ms, EMS memory time is +/- 6*5=30ms,
Delay time is 6*2 = 12ms. Delay Offset range: (-30+6) ~ (30-6) ms.

Returned Format:
The query returns the value of offset, and the unit is s.

Example:
:TIM:MODE DEL Set the scan mode as delayed scan.
:TIM:DEL:OFFS 1 Set the timebase offset as 1s.
:TIM:DEL:OFFS? Return 1.000e000.



4. :TIMebase[:MAIN]:SCALe

Command Format: