Command Systems

RIGOL

Programming Guide for DS1000B Series

2-67

:MEAS:DELASOUR?

Return CH1, CH2.



27. :MEASure:PHAseSOURce

Command Format:
:MEASure:PHAseSOURce ,
:MEASure:PHAseSOURce?

Functions
The command selects the channel for measuring the phase delay. The
may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.

Returned Format:
The query returns CH1, CH2 or CH1, CH3, CH1, CH4, CH2, CH3, CH2, CH4, CH3,
CH4.

Example:
:MEAS:PHASOUR CHAN1 CHAN2 Measure the phase delay.
:MEAS:PHASOUR?

Return CH1, CH2.


28. :MEASure:ENABle

Command Format:
:MEASure:ENAble

Functions
This command is open the locked MEASURE button and allow user to open Auto
Measure.

NOTE: Only corresponding unlock command can open AUTO and Auto Measure
once they are locked, neither restart nor update is available.

29. :MEASure:DISable

Command Format:
:MEASure:DISable

Function: