Model PD6000 Analog Input Process Meter Instruction Manual

80

Meter Copy or Cloning Instructions

!

Caution!

Do not connect the two meters to
the same signal source while
cloning. Internal calibration may
be affected.


1. Connect two meters using a PDA1200 meter copy cable.

Warning!

Using standard CAT5 or other cable will cause damage to
both meters. Use PDA1200 meter copy cable only.

2. Do not connect the two meters to the same signal source.
3. Power up both meters. Leave Clone meter in Run Mode.
4. Enter

the

Advanced

Features menu of the Master meter; see Ad-

vanced Features Menu on page 66.

5. Scroll to the Copy function using the Up arrow button then press

Enter.

6. The meter displays the message Send. Press Enter, the display

flashes while sending data. The message done is displayed when
copying is completed.

Send

Copy

done

Copy

Display flashes

while sending

data

Copy

7. The Clone meter displays the message CPY rC while being pro-

grammed then the message done when copying is completed. The
meter initializes and returns to Run Mode using the same settings
as the Master.

8. If meter to be cloned does not respond to the data being sent, refer