Meter Copy or Cloning Instructions
!
Caution!
Do not connect the two meters to 
the same signal source while 
cloning. Internal calibration may 
be affected. 
 
1. Connect two meters using a PDA1200 meter copy cable. 
Warning!
Using standard CAT5 or other cable will cause damage to 
both meters. Use PDA1200 meter copy cable only. 
2. Do not connect the two meters to the same signal source. 
3. Power up both meters. Leave Clone meter in Run Mode. 
4. Enter 
the
Advanced
Features menu of the Master meter; see Ad-
vanced Features Menu on page 66.
5. Scroll to the Copy function using the Up arrow button then press
Enter.
6. The meter displays the message Send. Press Enter, the display
flashes while sending data. The message done is displayed when 
copying is completed. 
 Send
 
Copy
 done
 
Copy
Display flashes
while sending
data
Copy
7. The Clone meter displays the message CPY rC while being pro-
grammed then the message done when copying is completed. The 
meter initializes and returns to Run Mode using the same settings 
as the Master. 
8. If meter to be cloned does not respond to the data being sent, refer
to Copy Function Requirements above.