FREQUENCY (DUTY CYCLE) MEASUREMENTS (ELECTRONIC)

1. Set the rotary function switch to the green Hz/% position.

2. Insert the black lead banana plug into the

negative COM jack and the red test lead
banana plug into the positive Hz jack.

3. Touch the test probe tips to the circuit under

test.

4. Read the frequency on the display.
5. Press the MODE button to indicate “%”.
6. Read the % duty cycle in the display.



% 4 – 20mA MEASUREMENTS

1. Set up and connect as described for DC mA measurements.
2. Set the rotary function switch to the 4-20mA% position.

3. The meter will display loop current as a % with 0mA=-25%,

4mA=0%, 20mA=100%, and 24mA=125%.

INSULATION RESISTANCE MEASUREMENTS

a)Set the rotary function switch to the INSULATION position

and Press the RANGE button to chose one of the voltages

displayed in the top left corner.

b)Connect two testing lines to the item to be tested.

c)Push down and hold the “TEST” button or press the “LOCK” key first

and then the “TEST” button, if there is any static electricity present

in the item being tested and its voltage is over 30V, it will fail and no

high-voltage testing will occur, “>30V” will display on the LCD, the

symbol “ ” flashes, and the buzzer will sound. Otherwise it will

enter into the formal testing process and the high-voltage will show

on the primary display, the insulation resistance in M

Ω is indicated

in-phase with the analog bar; on the top right corner display, the