2 Installation and Specifications

2.3 Sof

tw

a

re

set

u

p

Position the standard artifact along the R axis.

Align the artifact as closely as possible to the axis and then perform
a skew alignment as described in chapter 1 (see "Aligning the part
to an axis" on page 27).

Perform a single measurement of the entire range of motion using
the standard artifact and make a note of the result.

In this example of applying LEC, one point at end of the axis
measurement range is measured using an 8 inch standard.

Use an artifact that allows measurements of as much of
the axis range of motion as possible.

LEC example using an 8 inch standard

0

0

+0.02
+0.01

0.00

-0.01

-0.02

8.0

7.98

1

2

3

4

Arrow number

Descriptions

1

: Standard length

Entire 8 inch length is measured

2

: Standard values

The certified length of the standard

3

: Observed values

The measured length of the standard

4

: Deviation graph

Difference between standard and observed
values (not entered into any screen)