11451 Belcher Road South, Largo, FL 33773 • USA • Tel +1 (727) 447-6140 • Fax (727) 442-5699 • [email protected]
F-2600 & F-2700 Vortex Flow Meter Manual 02/15 - 0808-7 / 19204
Page 69
3.3.12 Calibration Menu
Password
ENTER
Use
keys to access menus.
Calibration
Menu
Meter Size
or Pipe ID
ENTER
Run Mode
Meter Factor
xxxx
Vortex Coef Ck
xx
Low Flow Cutoff
xx
Serial Number
xxxxxxxxx
F-2700 Series – meter size
F-2600 Series – pipe internal diameter (inches)
Meter calibration constant
– pulses / ft
3
– pulses / ft
Adaptive filter setting
Low Flow Cutoff
setting displayed
in volumetric flow
units (view only)
< Vol (xxx/xxx) >
xxx
< Mass (xxx/xxx)
xxx
Low Flow Cutoff
setting displayed
in mass flow
units (view only)
F-2700 Series
F-2600 Series
The Calibration Menu contains the calibration coefficients for the flow meter. These values
should be changed only by properly trained personnel. The Vortex Coef Ck and Low Flow
Cutoff are set at the factory. Consult the factory for help with these settings if the meter is
showing erratic flow rate.