8 fluid menu – ONICON F-2600 User Manual
Page 65
11451 Belcher Road South, Largo, FL 33773 • USA • Tel +1 (727) 447-6140 • Fax (727) 442-5699 • [email protected]
F-2600 & F-2700 Vortex Flow Meter Manual 02/15 - 0808-7 / 19204
Page 65
3.3.8 Fluid Menu
Password
ENTER
Use
keys to access menus.
Fluid
Menu
Flowing Fluid
Liquids >
Goyal-Dorais >
API 2540 >
Nat Gas AGA8 >
Real Gas >
Other Gas >
ENTER
Run Mode
< Liquid
Water
Ammonia
Chlorine
STD Press (PSIA)
xxxx
NORM Temp (C)
xxxx
NORM Press (KPA)
xxxx
STD Temp (F)
xxxx
Other Liquids >
< Mole Weight >
xxxx
< CRIT PRESS >
xxxx
< CRIT TEMP >
xxxx
< CRIT Z >
xxxx
< AL >
xxxx
< BL
xxxx
< Density @ 60F >
xxxx
< K0 >
xxxx
< K1 >
xxxx
< AL >
xxxx
< BL
xxxx
< Real Gas
Steam T & P Comp
< Specific Gravity >
xxxx
< MoleFract N2 >
xxxx
xxxx
xxxx
< Ref Press(PSIA)
xxxx
< Specific Gravity >
xxxx
< Compress (Z) >
xxxx
< Viscosity
xxxx
Air
Argon
Ammonia
CO
CO2
Helium
Hydrogen
Methane
Nitrogen
Oxygen
< Density >
xxxx
< AL >
xxxx
< BL
xxxx
Select “Steam T & P Comp” for VT and
in the run mode screens.
“Sat Steam T Comp” for the fluid type
VTP models. The VT model will display
For a V model in any fluid, enter nominal operating temperature and
pressure as simulated values in the diagnostics menu.
Liquified Gas >
< Liquified Gas
Carbon Dioxide
Nitrogen
Hydrogen
Oxygen
Nitrous Oxide
Argon
Propane
Use the Fluid Menu to configure the flow meter for use with common gases, liquids and
steam. Your flow meter is pre-programmed at the factory for your application’s process
fluid.
Reference Richard W. Miller, Flow Measurement Engineering Handbook (Third Edition,
1996), page 2-75 for definition and use of the Goyal-Doraiswamy equation and page 2-76 for
the definition and use of the API 2540 equation. Also, see Appendix B for Fluid Calculation
equations.
The units of measurement used in the Fluid Menu are preset and are as follows:
Mole Weight = lbm/(lbm•mol), CRIT PRESS = psia, CRIT TEMP = °R, Density = kg /m3 and
Viscosity = cP (centipoise).