F-2600 & F-2700 Vortex Flow Meter Manual 02/15 - 0808-7 / 19204
Page A-7
Output Signals
(1)
Analog: volumetric meter-field rangeable linear 4-20 mA output signal (1200
ohms maximum loop resistance) selected by user for mass flow rate or
volumetric
flow
rate.
Communications: HART, MODBUS-RTU, RS485, BACnet MS/TP
Multiparameter Meter: Up to three field rangeable linear 4-20 mA output
signals (1200 ohms maximum loop resistance) selected from the five
parameters mass flow rate, volumetric flow rate, temperature, pressure and
density.
Pulse: pulse output for totalization is a 50-millisecond duration pulse
operating a solid-state relay capable of switching 40 VDC, 40 mA maximum.
Note: (1) All outputs are optically isolated and require external power for
operation.
Alarms
Up to three programmable solid-state relays for high, low or window alarms
capable of switching 40 VDC, 40 mA maximum.
Totalizer
Based on user-determined flow units, six significant figures in scientific
notation. Total stored in non-volatile memory.
Wetted Materials
In-Line Flow Meter:
316L
stainless
steel
standard
C276 hastelloy or A105 carbon steel optional
Insertion
Flow
Meter:
316L
stainless
steel
standard
Teflon® packing gland below 500° F (260° C)
Graphite packing gland above 500° F (260° C)
Enclosure Protection
NEMA 4X and IP66 cast enclosure
Classification
Electrical Ports
Two 3/4" female NPT ports
Mounting Connections
In-Line: Wafer, 150, 300, 600 lb ANSI flange, PN16, PN40, PN64 flange
Insertion Hot Tap(1) Installation: 2" MNPT; 150, 300, 600 lb
ANSI flange, PN16, PN40, PN64 flange and optional retractor
with packing gland probe seal.
Note: (1) Removable under line pressure.
Mounting Position
In-Line Flow Meter: No effect.
Insertion Flow Meter: Meter must be perpendicular within ± 5° of the pipe
centerline.
Certifications
Material Certificate – US Mill certs on all wetted parts
Pressure
Test
Certificate
Certificate
of
Conformance
NACE
Certification
(MR0175)
Oxygen
Cleaning
(CGA
G-4.)