beautypg.com

ONICON F-2600 User Manual

Page 104

background image

11451 Belcher Road South, Largo, FL 33773 • USA • Tel +1 (727) 447-6140 • Fax (727) 442-5699 • [email protected]

F-2600 & F-2700 Vortex Flow Meter Manual 02/15 - 0808-7 / 19204

Page 104

5.3 TROUBLESHOOTING THE FLOW METER

5.4 FIRST CHECK ITEMS:

Installation Direction Correct

Installation Depth Correct (Insertion style meter)

Power and Wiring Correct

Application Fluid Correct

Meter Range Correct for the Application

Meter Configuration Correct

Describe Installation Geometry; i.e. upstream diameters, valve position,

downstream diameters, etc.

5.5 RECORD VALUES:

Record the following values from the Run Menu with the meter installed in order to
determine the operating state of the flow meter:

With Flow

With No Flow (if possible)

Flow =

Temperature=

Pressure =

Density =

Error Messages? =

Record the following values from the Hidden Diagnostics Menu with the meter
installed:
(Use password 16363 to access.)

With Flow

With No Flow (if possible)

f =

fi =

A =

A1 =
A2 =
A3 =
A4 =

V =

RTD1 =
RTD2 =

!

WARNING

Before attempting any flow meter repair, verify that the line is not pressurized. Always remove

main power before disassembling any part of the mass flow meter. Use hazardous area precautions

if applicable. Static sensitive electronics - use electro-static discharge precautions.

This manual is related to the following products: