5.3 TROUBLESHOOTING THE FLOW METER
5.4 FIRST CHECK ITEMS:
□
Installation Direction Correct
□
Installation Depth Correct (Insertion style meter)
□
Power and Wiring Correct
□
Application Fluid Correct
□
Meter Range Correct for the Application
□
Meter Configuration Correct
□
Describe Installation Geometry; i.e. upstream diameters, valve position,
downstream diameters, etc.
5.5 RECORD VALUES:
Record the following values from the Run Menu with the meter installed in order to
determine the operating state of the flow meter:
With Flow
With No Flow (if possible)
Flow =
Temperature=
Pressure =
Density =
Error Messages? =
Record the following values from the Hidden Diagnostics Menu with the meter
installed: (Use password 16363 to access.)
With Flow
With No Flow (if possible)
f =
fi =
A =
A1 =
A2 =
A3 =
A4 =
V =
RTD1 =
RTD2 =
!
WARNING
Before attempting any flow meter repair, verify that the line is not pressurized. Always remove
main power before disassembling any part of the mass flow meter. Use hazardous area precautions
if applicable. Static sensitive electronics - use electro-static discharge precautions.