Weight
Flow Meter Size
L
H
*
Ø
½" (15 mm)
2.6 in
(65 mm)
13.3 in
(338 mm)
1.38 in
(35 mm)
11.5 Lb
(5.2 Kg)
¾" (20 mm)
2.6 in
(65 mm)
13.4 in
(340 mm)
1.69 in
(42.9 mm)
11.9 Lb
(5.4 Kg)
1" (25 mm)
2.6 in
(65 mm)
13.5 in
(343 mm)
2.00 in
(50.8 mm)
12.2 Lb
(5.5 Kg)
1½" (40 mm)
2.6 in
(65 mm)
13.8 in
(351 mm)
2.88 in
(73.2 mm)
13.8 Lb
(6.3 Kg)
2" (50 mm)
2.6 in
(65 mm)
14.0 in
(356 mm)
3.62 in
(91.9 mm)
15.5 Lb
(7.0 Kg)
3" (80 mm)
2.6 in
(65 mm)
14.5 in
(368 mm)
5.00 in
(127.0 mm)
20.6 Lb
(9.4 Kg)
4" (100 mm)
2.6 in
(65 mm)
15.0 in
(381 mm)
6.19 in
(157.0 mm)
25.3 Lb
(11.5 Kg)
*
This Dimension is the same for integral and remote mount meters.
Power Requirements 12 to 36 VDC, 25 mA, 1 W max., loop-powered volumetric or, mass
12 to 36 VDC, 300 mA, 9 W max. multiparameter mass options
100 to 240 VAC, 50/60 Hz, 5 W max. multiparameter Mass options
Class I Equipment (grounded type)
Installation (Over-voltage) Category II for transient over-voltages
AC & DC mains supply voltage fluctuations are not to exceed
±10% of the rated supply voltage range.
User is responsible for the provision of an external Disconnect Means (and
Over-Current Protection) for the equipment (both AC and DC models).
Display
Alphanumeric 2 x 16 LCD digital display
Six push-button switches (up, down, right, left, enter, exit) operable through
explosion-proof window using hand-held magnet. Viewing at 90o mounting
intervals.
Pocess Fluid and
Process Fluid:
Ambient Temperature Standard temperature sensor: –330° F to 500° F (–200° C to 260° C)
High temperature sensor: to 750° F (400 °C)
Ambient:
Operating temperature range: –40° F to 140° F (–40° C to 60° C)
Storage temperature range: –40° F to 185° F (–40° C to 85° C)
Maximum relative humidity: 0-98%, non-condensing conditions
Maximum altitude: -2000 to 14,000 feet (-610 to 4268 meters)
Pollution Degree 2 for the ambient environment