beautypg.com

MK Products CobraTig 150 XM User Manual

Page 26

background image

CobraTig 150 Owner's Manual P/N 091-0417 - Page 26

This screen will allow the operator to calibrate any orbital weldhead,

at any speed value, between 1.0 and 10.0 rpm. Using the HIGH/LOW

button, select the respective speed value and use the ADJUST UP and

ADJUST DOWN buttons to select the rpm value required.

The HIGH and LOW values can never be equal. A minimum of 1.0 rpm

will separate the two values. If the LOW value is increased, the HIGH

will increase, with the 1.0 rpm difference between them. Similarly, if the

HIGH is decreased, the LOW will decrease with the 1.0 rpm difference.

Feedback of the weldhead during calibration is seen in the bar graph on

the left side of this screen.

High/Low

Toggles between the HIGH and LOW rpm speed values.

Print Report

Press this button, once calibration is complete, to view a printed report of

the calibration values.

The calibration values shown are an indication of the MOTOR/TACH

status of the weldhead. A typical printout is shown below:

MOTOR Calibration report

MOTOR gain: -4.6% Zero offset:

0.30 RPM

TACH gain:

1.9% Zero offset:

0.02 RPM

Should these MOTOR & TACH values return with number of 0.00% and

0.00 RPM, the calibration was invalid and should be re-run. Continu-

ous results of all zeros, is an indication that the weldhead is in need of

service.

Consult the weldhead owner’s manual for proper maintenance and cali-

bration service.

MOTOR Calibration report

MOTOR gain: 0.0% Zero offset:

0.00 RPM

TACH gain:

0.0% Zero offset:

0.00 RPM

NOTE: If all values for MOTOR or WELDER calibration are zeros, it

means that he calibration did not complete successfully and is not valid.

Start Calibration

Press this button to start the weldhead calibration procedure. The cali-

bration start at the values selected using the “ADJUST UP” and “AD-

JUST DOWN” buttons.

Current Start

Using the calibration kit and the multimeter, calibration of the output cur-

rent can be achieved in relatively short time. Using the 10 A CALIBRA-

TION and 80 A CALIBRATION buttons, output values at these points will

properly calibrate the current for almost all applications.

10A Calibration / 80A Calibration

These two keys are used in conjunction for calibrating the output cur-

rent. Use the information on the HELP screen to perform this calibration

procedure.

This manual is related to the following products: