CobraTig 150 Owner's Manual P/N 091-0541 - Page 25

Options from the CALIBRATION MENU include motor output and tachometer

feedback, current output and feedback, and arc voltage input.

In order to properly calibrate this machine, you will need the following: DC

power supply capable of producing 1.0 VDC to 20.0 VDC, a certified digital

multimeter, and various standard connection leads. To aid in the calibration

of this machine, purchase kit P/N 005-0251 (The instructions included with

the kit are for the model ACL Orbital Welder, they do not apply to this unit).

Use the instructions on the HELP screen for a full explanation on how to

calibrate this unit.

Help

Press this button to read a brief description of the CALIBRATION MENU

button descriptions and functions.

Main Menu

Press this button to return to the MAIN MENU.

1.8 VDC Calibration / 18 VDC Calibration

These two keys are used in conjunction for calibrating the arc voltage

feedback circuit. Use the information on the HELP screen to perform this

calibration procedure.

Connect an external DC voltage power supply to the shunt connectors

labeled ‘ARC VOLTAGE’ and adjust to 1.8 VDC input. Reference the

VDC bar graph on the LCD screen, and use the ADJUST UP or ADJUST

DOWN buttons to match bar graph to input value. Once bar graph value

is matched to the input value, press the 1.8 VDC button, this will set the

calibration.

Set the input to 18.0 VDC and repeat the above process. The procedure

will cycle through twice for an average reading and setting, then quit

when ready.

Adjust Up

Press this button to adjust the bar graph setting to match that of the ref-

erence value being read - voltage or current.

Motor Calibration

With the weldhead plugged into the CobraTig® 150 XM rear panel, the

motor voltage output to the weldhead, and tachometer feedback from the

weldhead rotor can be calibrated.