Transmit parameters to the sensor – Microsonic LCA-2 User Manual
Page 13

Reading / writing parameters
Page 13 of 51
T
RANSMIT PARAMETERS TO THE SENSOR
All changes that you have made in the input masks, are only temporarily stored within the
LinkControl program. You have to transfer the data to the sensor and / or save it as a file on the hard disc
of your PC or on a floppy disc.
Click
for sending the parameters to the sensor
Confirm the security query
It is checked in advance whether the connected sensor matches the sensor type, which is indicated in the
in-put mask. Subsequently the altered parameters are transferred to the sensor.
If the following message occurs,
push the RESET button on the LinkControl adaptor.
If still the parameters cannot be written to the sensor, please check, whether
a sensor capable for LinkControl is connected
the sensor gets its power supply via the LinkControl adapter (LED's on sensor have to be
illuminated)
the USB connection to the PC is properly made
the correct port has been selected within the LinkControl software
Also: Reading parameters from a connected sensor
If the LinkControl software realises, that the selected sensor type does not match the connected sensor
type, the transfer of parameters is inhibited.