beautypg.com

Xfrsize, Aitrig, Ranges – Measurement Computing DAQFlex User Guide User Manual

Page 80: Rearm

background image

DAQFlex Software User's Guide

DAQFlex Message Reference – Reflection messages – AITRIG

80

XFRSIZE

Get the number of bytes used in the transfer of each data sample.
Message

"@AISCAN:XFRSIZE"

Response

"AISCAN:XFRSIZE=<implementation>value"

implementation FIXED%, or not specified if value is NOT_SUPPORTED

value

The number of bytes, or NOT_SUPPORTED

Example

"AISCAN:XFRSIZE=FIXED%2"

AITRIG

Get the analog input trigger properties of a device.

Properties

RANGES, REARM, SRCS, TYPES

RANGES

Get the supported ranges for a device's analog input trigger circuit.
Message

"@AITRIG:RANGES"

Response

"AITRIG:RANGES=implementation>value"

implementation FIXED%, PROG% (programmable), HWSEL% (hardware selectable),

or not specified if value is NOT_SUPPORTED

value

A list of all valid ranges for the specified analog trigger source or

device, or returns NOT_SUPPORTED if the device doesn't support

analog triggering of analog input or the trigger source is not valid for

the device.

Example

"AITRIG:RANGES=PROG%BIP10V,BIP5V"

Note

On some devices, the values returned may be dependent on channel configuration
settings.

REARM

Get a value indicating whether the device supports continuous triggering of the analog input.
Message

"@AITRIG:REARM"

Response

"AITRIG:REARM=implementation>value"

implementation FIXED%, PROG%, or not specified if value is NOT_SUPPORTED

value

ENABLE, DISABLE, or returns NOT_SUPPORTED if the device doesn't

support continuous triggering of analog input.

Example

"AITRIG:REARM=PROG%ENABLE,DISABLE"