Measurement Computing DAQFlex User Guide User Manual
Page 109
DAQFlex Software User's Guide
Hardware Reference – USB-1608FS-Plus
109
Component
Supported
Property/Command
Set/Get
Supported Values
Get
Number of counter channels
VALUE
Get
0 to 4,294,967,295
Set
0
FLASHLED/{n}
Set
0 to 255
FWV
Get
Firmware version of the device
MM.mm (M = major; m = minor)
ID
Set/Get
Up to 56 characters
MFGCAL
Get
yyyy-mm-dd HH:MM:SS
MFGCAL{item}
Get
YEAR as yyyy; 20xx
MONTH as mm; 01 to 12
DAY as dd; 01 to 31
HOUR as HH; 01 to 23
MINUTE as MM; 01 to 59
SECOND as SS; 01 to 59
MFGSER
Get
Up to 8 hexadecimal digits
RESET/{value}
Set
DEFAULT
Get
Number of digital ports
Get
1
DIR
Set/Get
IN, OUT
LATCH
Set/Get
0 to 255
VALUE
Set/Get
0 to 255
DIR
Set/Get
IN, OUT
LATCH
Set/Get
port number: 0
bit number: 0 to 7
port value: 0 to 255
bit value: 0, 1
VALUE
Set/Get
port number: 0
bit number: 0 to 7
port value: 0 to 255
bit value: 0, 1
Hardware features
8 single-ended 16-bit simultaneous analog input channels, numbered 0 to 7.
o Analog voltage input ranges:
±10V
±5V
±2V
±1V
±0.625V
±0.3125V
o Sampling rate:
100 kS/s max rate for one channel
400 kS/s max aggregate rate, streaming. Limited by USB transfer speed; 800 kS/s max to 32K
sample FIFO
o Channel-gain queue– eight elements; one gain element per channel.
8 DIO, individually configurable